EasyManuals Logo
Home>Honeywell>Security Sensors>XT Series

Honeywell XT Series User Manual

Honeywell XT Series
28 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #5 background imageLoading...
Page #5 background image
XT Series Personal Gas Detector
5
Warning: Read Manual Prior to Use
1. Introduction
The XT is a disposable, easy to use, personal gas detector, designed for 2 years continuous
monitoring of the atmosphere for potentially hazardous levels of oxygen deficiency or toxic gas.
SAFETY INFORMATION
Substitution of components may impair intrinsic safety•
Do not activate the detector after the date on the packaging•
Perform a self-test prior to each day’s use•
Do not use in oxygen-enriched atmospheres•
Periodically test the sensors response to gas by exposing the detector to a target gas •
concentration that exceeds the alarm set points. Manually verify that the audible, visual
and vibrating alarms are activated
Exposure to extremely high levels of over-range gas may cause temporary loss of sensor •
accuracy. Allow monitor to stabilize in zero gas atmosphere before reuse. It is recommended
to perform sensor response gas test prior to reuse.
IMPORTANT NOTICE
Honeywell Analytics can take no responsibility for use of its equipment if it is not done in accordance
with the appropriate issue and/or amendment of the relevant manual. If further details are required
which do not appear in this manual, contact Honeywell Analytics or their agent.
Honeywell Analytics shall not be liable for any incidental or consequential damages in connection
with any deletions, errors or omissions in this Manual.

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Honeywell XT Series and is the answer not in the manual?

Honeywell XT Series Specifications

General IconGeneral
BrandHoneywell
ModelXT Series
CategorySecurity Sensors
LanguageEnglish

Related product manuals