40 tina83e1-d (2018-12)
For adjusting the zero, operate the gauge under the same con-
stant ambient conditions and in the same mounting orientation
as normally.
The zero can be adjusted via
the <ADJ> button on the gauge,
the diagnostic port ( [1]),
the RS232C/485C interface ( [1]),
the EtherCAT interface ( [2]).
If you are using a seal with centering ring and filter, check
that they are clean or replace them if necessary
( "Deinstallation").
Put the gauge into operation and operate it at atmospheric
pressure for at least 10 minutes.
Briefly press the <ADJ> button with a pin (max. ø1.1 mm)
and the ATM adjustment is carried out, or
ATM
HV
perform the adjustment via the diagnostic port or via the
EtherCAT interface.
The Pirani sensor is adjusted to 1000 mbar (duration 5 s).