Power Supply Function
4.4.2 Set the Internal Resistance
The IT6000C series power supply provides internal resistance setting (CV prior-
ity mode only). The procedures are shown as below.
1. Press the composite keys [Shift]+[V-set] (Config) on the front panel to enter
the configuration menu.
2. Press the up/down key or rotate the knob to select Output Res and press
[Enter] to confirm.
3. Set the internal resistance value and press [Enter] to confirm.
4.4.3 Set the Output-On/Output-Off Delay
You can set the output-on/output-off delay time within the range from 0 seconds
to 60 seconds.
• On Delay: Indicates from the time that a command to turn on the output is re-
ceived until the output actually turns on.
• Off Delay: Indicates from the time that a command to turn off the output is re-
ceived until the output actually turns off.
The procedures to set the output delay time are as follows.
1. Press the composite keys [Shift]+[V-set] (Config) on the front panel to enter
the configuration menu.
2. Press the up/down key or rotate the knob to select On Delay or Off Delay,
and press [Enter] to confirm.
3. Set the output-on/output-off delay time and press [Enter] to confirm.
4.5 Protection Function for Power Supply
The IT6000C series power supply provides the general protection functions
such as overvoltage, overcurrent, overpower, undercurrent and undervoltage
protection. The corresponding protection parameters can be configured in the
Protect menu. In addition, the power supply also provides over temperature pro-
tection and Sense reverse connection protection.
The descriptions of Protect menu are listed in the table below.
Protect
Protection function menu for the power supply
OVP
Overvoltage protection
Off
Turn the OVP function off. (Def)
On Turn the OVP function on.
Copyright © Itech Electronic Co., Ltd.
51