EasyManuals Logo
Home>Kane>Measuring Instruments>KM940

Kane KM940 User Manual

Kane KM940
35 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #9 background imageLoading...
Page #9 background image
- 9 -
4. NORMAL START UP SEQUENCE
4.1 Every Time You Use The Analyser
BEFORE SWITCH-ON CHECK THAT:
the particle filter is not dirty
the water trap and probe line are empty of water
all hose connections, etc, are properly made
the probe is sampling CLEAN AMBIENT air
the water trap is correctly fitted and the instrument upright
the flue temperature is connected
Switch ON the instrument by pressing
4.2 Automatic Calibration
During this sequence the analyser pumps fresh air into the sensors to allow toxic sensors (if
fitted) to be set to zero and the Oxygen sensor to be set to 20.9 %.
After switch-on the analyser will briefly display header information :-
Kane International
(44)-1707-375550
And then show the countdown screen :-
ZERO CAL
Time : 180
FRESH AIR PURGE
The calibration time will count down in seconds to zero. Calibration time may be changed
from 2 to 6 minutes. See Set-Up menu section 5.2.5.
Note ! Three minutes is recommended to allow the sensors to stabilise fully. Anything less
than this may result in drift of the toxic and oxygen sensors in clean ambient air.
To obtain the quoted specification an instrument should be calibrated with clean ambient air
at standard temperature and pressure (STP).

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Kane KM940 and is the answer not in the manual?

Kane KM940 Specifications

General IconGeneral
BrandKane
ModelKM940
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals