Specifications
Table1: Series2200specifications (cont.)
Parameter Description
Current resolution,
nominal
1mA
Current set
ting
accuracy
±(0.1% + 5 mA) at 25 °C ±5 °C
Voltage li
mit,
nominal
0to30V
Current
temperature
coefficie
nt, typical
±(0.1% + 5 mA) per °C outside the 25 °C ± 5 °C standard range
Meter, nominal Internal readback capability. Voltage and current can be measured internally.
Current readback
resolut
ion, nominal
1mA
Current
readback accuracy
±(0.1%
of reading + 5 mA) at 25 °C ±5 °C
Current readback
temperature
coefficient, typical
±(0.1% + 5 mA) per °C outside of the 25 °C ±5 °C standard range
Voltage
readback accuracy
±(0.03% of reading + 10 mV) at 25 °C ±5 °C standard range
Voltage readback
resolution, nominal
1mV
Voltage readback
temperature
coef
ficient, typical
±(0.03% of reading + 10 mV) per °C outside the 25 °C ±5 °C standard range
Line regulation
– voltage
Chan
ge over the full rated line voltage range: ±(0.01% + 3 mV)
Line regulation
–cur
rent
Change over full rated line voltage range: ±(0.1% + 3 mA)
Load regulation
– voltage
Cha
nge from 0 to 100% of full scale current: ±(0.01% + 3 mV)
Load regulation
–cu
rrent
Change from 10 to 90% of full scale voltage: ±(0.01% + 3 mA)
Voltage ripple,
20 MHz bandwidth
<3mV
RMS
and 20 mV
p-p
<1.
5mV
RMS
and 12 mV
p-p
,typical
Voltage ripple,
7M
Hz bandwidth
<1mV
RMS
and 3 mV
p-p
Current ripple,
20 MHz bandwidth
<5mA
RMS
Common Mode
Current Noise
<1.5μA
RMS
, 700 Hz bandwidth.
2 Series 2200 Multichannel Programmable DC Power Supplies Technical Reference