EasyManuals Logo
Home>Keithley>Power Supply>2200 Series

Keithley 2200 Series User Manual

Keithley 2200 Series
66 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #16 background imageLoading...
Page #16 background image
Specications
Table 3: 2230-30-1 and 2230J-30-1 specic specications for Channel 3 (cont.)
Constant Current Operation,
nominal
The unit may be set to be a constant current over a range of voltages.
Current Range The output voltage may be adjusted from 0 A to 5 A.
Current Resolution, nominal
1mA
Current Setting Accuracy
+/-(0.1% + 5 mA) at 25 ° C ±5 °C.
Voltage Limit, nominal 0 to 6 V.
Current Temperature Coefcient,
typical
+/-(0.1% + 5 mA) per °C outside of the 25 °C ±5 °C s tandard range
Meter, nominal Internal readback capability. Voltage and current can be m easured internally. DVM connections
exist to allow measuring external voltages.
Current Readback Resolution,
nominal
1mA
Current Readback Accuracy
±(0.1% of reading + 5 mA) at 25 °C ±5 °C
Current Readback Temperature
Coef
cient, typical
±(0.1% +5mA) per °C outside of the 25 °C ± 5 °C standard range
Voltage Readback Accuracy
±(0.0
3% of reading + 10 mV) at 25 °C ±5 °C
Voltage Readback Resolution,
nominal
0.1 mV
Voltage Readback Temperature
Coefcient, typical
±(0.03% of reading + 10 mV) per °C outside of the 2 5 °C ±5 °C standard range
Line Regulation Voltage
Change over the full rated line voltage range: ±(0.01% + 3 mV)
Line Regulation Current
Change over full rated line voltage range: ±(0.1% + 3 mA)
Load Regulation Voltage
Change from 0 to 100% of full scale. Voltage: ±(0.01% + 3 mV)
Load Regulation Current
Change from 10 to 90% of full scale. Current: ± (0.01% + 3 mA)
Volta
ge Ripple, 20 MHz
Bandwidthh
<3mV
RMS
and 20 mV
p-p
.
<1.5mV
RMS
and 12 mV
p-p
,typical
Voltage Ripple, 7 MHz
bandwidth
<1mV
RMS
and3mV
p-p
Curre
nt Ripple, 20 MHz
Bandwidth
Less than 6 mA
RMS
Common Mode Current Noise
<1.5
μA
RMS
. 700 Hz bandwidth.
Volt
age Transient Response
Settling Time, Load Change,
typical
<150 ms to within 75 mV following a change from 0.1 A to 1 A.
Voltage Transient Response
Settling Time, Setting Change,
Ris
ing, typical
<150 ms from beginning of excursion to within 75 mV of terminal value following a change
fro
m0.4Vto4Vintoa4 resistor.
Voltage Transient Response
Se
ttling Time, Setting Change,
Falling, typical
<1
50 ms from beginning of excursion to within 75 m V of terminal value following a change
from0.4Vto4Vintoa4 resistor.
The specication does not include Command Decode Time.
6 Series 2200 Multichannel Programmable DC Power Supplies Technical Reference

Other manuals for Keithley 2200 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2200 Series and is the answer not in the manual?

Keithley 2200 Series Specifications

General IconGeneral
BrandKeithley
Model2200 Series
CategoryPower Supply
LanguageEnglish

Related product manuals