Specications
Table 10: Mecha
nical characteristics
Parameter
Weight,
instrument,
nominal
9kg
(19.8 lbs)
Model
Dimensions (Width * Height * Depth
Dimensions,
nominal
2230
214 mm * 88 mm * 355 mm (8.45” * 3.47” * 13.9”)
Cooling method,
nominal
Fan
Required cooling
clearance
Rear of instrument must be separated by at least 5 cm (2 in) from any airow restrictions to assure adequate
cooling.
Bottom of instrument must have 1 cm (0.4 in) from any airow restrictions. The protective boot maintains
the proper bottom clearance.
For rackm
ount conguration, follow the clearance requirements provided with the rackmount installation
instructions.
Construction
materials, nominal
Chassis Aluminum
Front Panel Plastics PC/A BS blend
Table 11: Environment performance
Parameter All m od els
Temperature
Operating: +0 °C to +40 °C
Nonoperating: -20 °C to 70 °C
Humidity
Operating:
5% to 95% relative humidity (% RH) at up to 40 °C, noncondensing
Nonoperating:
5% to 95% RH (relative humidity) at up to +40 °C,
5% to 60% RH above +40 °C up to +70 C, noncondensing
Altitude
Operating: 100% capability up to 2,000 meters.
Nonoperating: Up to 4,000 meters
Table 12: Keypad special operations
Ch
aracteristic
De
scription
Display the heat sink temperature Enter + Esc + 1
Reset the unit to the factory default
s
ettings
Enter+Esc+2
10 Series 2200 Multichannel Programmable DC Power Supplies Technical Reference

Other manuals for Keithley 2220-30-1

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2220-30-1 and is the answer not in the manual?

Keithley 2220-30-1 Specifications

General IconGeneral
BrandKeithley
Model2220-30-1
CategoryPower Supply
LanguageEnglish

Related product manuals