Specifications
Table 13: Safet
y characteristics
Item Description
U.S. National
ly Recognized Testing Laboratory (NRTL) Listing
UL61010-1–2004 Safety Requirements for Electrical Equipment for Measurement, Control,
and Laboratory Use
Canadian Certification
CAN/CSA C22
.2 No. 61010-1-2004 Safety R equirements for Electrical Equipment for
Measurement, Control, and Laboratory Use
European Union C om pliance
Low Voltage Directive 2006/95/EC
EN61010-1
2001 Safety Requirements for Electrical Equipment for Measurement, Control,
and Laboratory Use
Safety Certifi
cations
Additional Compliance
IEC61010-1 2001 Safety Requirements for Elec trical Equipment for Measurement, Control,
and Labor
atory Use
Channel
to Channel Isolation
Voltage
Any outp
ut can be floated up to 240 V (DC + peak AC with AC limited to a maximum 3V
pk-pk
and
maximum of 60 Hz) relative to any other channel terminal. Remote sense terminals should
remain within 1 V of their respective outputs to avoid possible damage.
Channel to Ground Isolation
Voltag
e
Any output can be floated up to 240 V (DC + peak AC with AC limited to a m aximum 3V
pk-pk
and
maximu
m of 60 H z) relative to the Earth Ground terminal. Remote s ense terminals should
remain within 1 V of their respective outputs to avoid possible damage.
Series 2200 Multichannel Programmable DC Power Supplies Technical Reference 11