2260B Series Programming Manual
48
OUTPut:MODE
Description
Sets the 2260B output mode. This is the equivalent
to the F-03 (V-I Mode Slew Rate Select) settings.
Syntax
Return Syntax
OUTPut:MODE {<NR1>|CVHS|CCHS|CVLS|CCLS}
OUTPut:MODE?
Parameter 0
CV high speed priority
CVHS
CV high speed priority
1
CC high speed priority
CCHS
CC high speed priority
2
CV slew rate priority
CVLS
CV slew rate priority
3
CC slew rate priority
CCLS
CC slew rate priority
Return parameter <NR1>
Returns the output mode.
OUTPut[:STATe][:IMMediate]
Description
Turns the output on or off.
Syntax
Query Syntax
OUTPut[:STATe][:IMMediate] { OFF | ON | 0 | 1 }
OUTPut[:STATe][:IMMediate]?
Parameter 0
<NR1> Turns the output off.
OFF
Turns the output off.
1
<NR1> Turns the output on.
ON
Turns the output on.
Return parameter <NR1>
Returns output status of the instrument.
OUTPut[:STATe]:TRIGgered
Description
Turns the output on or off when a software trigger
is generated.
Syntax
Query Syntax
OUTPut[:STATe]:TRIGgered { OFF | ON | 0 | 1 }
OUTPut[:STATe]:TRIGgered?
Set
Query
Set
Query
Set
Query

Table of Contents

Other manuals for Keithley 2260B-30-72

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2260B-30-72 and is the answer not in the manual?

Keithley 2260B-30-72 Specifications

General IconGeneral
BrandKeithley
Model2260B-30-72
CategoryPower Supply
LanguageEnglish

Related product manuals