5: Verification test record forms Series 2260B Programmable Power Supplies
Verification and Adjustment Manual
5-2 077104802 / November 2016
30 Volt verification test record
Print this page and record your verification results that can be used for reference in the future. Make
sure to keep this record with your power supply.
Constant voltage tests:
Voltage programming
and measurement
Model Minimum
specifications
Results Maximum specifications
CV load regulation Model Minimum
specifications
Results Maximum specifications
CV line regulation Model Minimum
specifications
Results Maximum specifications
CV ripple and noise Model Minimum
specifications
Results Maximum specifications
Peak-to-peak
30-36 N/A 60 mV
30-72 N/A 80 mV
30-108 N/A 100 mV
RMS
30-36 N/A 7 mV
30-72 N/A 11 mV
30-108 N/A 14 mV