EasyManuals Logo
Home>Keithley>Measuring Instruments>SourceMeter 2410

Keithley SourceMeter 2410 Quick Results Guide

Keithley SourceMeter 2410
53 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #18 background imageLoading...
Page #18 background image
10 Quick Results Guide
Remote command programming
Table 1 lists the SCPI commands to select terminals, sense, and cable guard.
Basic SourceMeter operations
Fundamental SourceMeter operations include source-measure, measure only (V or I) and
measure ohms.
Source-measure
There are four source-measure combinations for the SourceMeter:
Source V, Measure I
Source I, Measure V
Source V, Measure V
Source I, Measure I
The basic procedure to source-measure is provided in Table 2. It assumes that the DUT is
already connected to the SourceMeter as explained in
Basic connections
.
Table 1
SCPI commands; terminals, sense, and cable guard
Command Description
:ROUTe:TERMinals <name> Selects terminals; <name> = FRONt or REAR.
:SYSTem:RSENse <b> Selects sense mode; <b> = ON (4-wire) or OFF (2-wire).
:SYSTem:GUARd <name> Selects guard mode; <name> = CABLe or OHMS.

Table of Contents

Other manuals for Keithley SourceMeter 2410

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley SourceMeter 2410 and is the answer not in the manual?

Keithley SourceMeter 2410 Specifications

General IconGeneral
BrandKeithley
ModelSourceMeter 2410
CategoryMeasuring Instruments
LanguageEnglish