7.3 Data output RS 232 C
Technical Data
8-bit ASCII Code
1 start bit, 8 data bits, 1 stop bit, no parity bit
Baud rate selectable from 2400, 4800, 9600 Baud (factory setting) and 19200 Baud.
Sub-D plug 9-channel required
For operation with interface faultless operation is only ensured with the correct KERN interface cable
(max. 2m)
Pin allocation of the balance output socket (front view)
Sub-D jack 9-channel
5 1
9 6
Pin 2: Transmit data
Pin 3: Receive data
Pin 5: Signal ground
Baud rate
The MODE key is used to set the Baud rate used for the transfer of measured values. In the following example
the baud rate is set to 4800 baud.
Setting Baud rate KERN CKE
Display
1. Switch on balance
2. Press and hold tare key
3. Tap ON/OFF key and let go of tare key
"Cal?" will appear on the display.
4. Keep holding the switch over key „pcs ↔ g“ until "Printer" appears on the
display and confirm by pressing the "20x" key. 2.400 Bd will appear on the
display.
5. Press the switchover key "pcs ↔ g" to select the Baud rate and confirm
by the “20x” key. To return to weighing mode, press the tare key.
PRINTER?
2400
Baud
4800
Baud
4800 Baud X
0.00 oz
CKE/CDS-BA-e-1526 18

Other manuals for KERN CKE 360-3

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the KERN CKE 360-3 and is the answer not in the manual?

KERN CKE 360-3 Specifications

General IconGeneral
BrandKERN
ModelCKE 360-3
CategoryScales
LanguageEnglish

Related product manuals