- 16 - EOB-B_EOE-B_EOS-B-SH-e-1911
Calibration procedure completed and the scale will return
back to weighing mode automatically
LF 2 Spec calibration
Press to enter
Confirm display of [ECF 1] with . [CAL Z] is displayed.
Internal value
Press to confirm
A : Metric Units 0 : None 1 : kg
B : American Units 0 : None 1 : lb
C : Fixed to 0
D : PCS 0 : Off 1 : On
E : Duo range 0 : Off 1 : Multi interval
F : Calibration Unit
1 : Metric units as calibration unit, single calibration
2 : American units as calibration unit, single calibration
3 : Metric units as calibration unit, two sections calibration
4 : American units as calibration unit, two sections calibration
The scale will not allow to continue to the next setup step
if an error has occurred during the programming
A B C D E F
Press to confirm
Switch to the next digit to the right with .
Increase number with .
Press to confirm