EW-BA-defsi-0620
29
12 Trouble shooting
No indication. The balance is not turned on.
The mains connection does not
function (the cable is not connected,
the cable is defective).
The mains supply has broken down.
The weighing display changes Air movements
constantly. Vibrations of the table or the floor.
The pan is in contact with an alien
element.
The adjusting is no longer correct.
The temperature varies strongly.
Error messages
o-Err The weighing range is exceeded.
b-Err Electronic error caused by
electrostatic field. Check
environment. If necessary, choose
another location.
u-Err The weighing pan does not lie on the
pan support.
1-Err Wrong adjusting weight (too light).
2-Err The adjusting weight is too incorrect.
3-Err At the beginning of the adjusting, a
weight is on the balance.
4-Err Grave error.
CAL. OFF Error message only with models
having an internal adjusting weight.
The adjusting switch (right) is not in
the “WEIGH“ position.
If other error messages occur, turn the balance off and on again. The weighing result
is obviously wrong.
If the error message still appears, please contact KERN.