EasyManuals Logo
Home>KERN>Scales>EWJ

KERN EWJ User Manual

KERN EWJ
77 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #76 background imageLoading...
Page #76 background image
76 TEWJ-B-BA-e-2310
17 Instant help for troubleshooting
In case of an error in the program process, briefly turn off the balance and disconnect
from power supply. The weighing process must then be restarted from the beginning.
Fault
Possible cause
The weight display does not
glow.
The balance is not switched on.
The mains supply connection has been interrupted
(mains cable not plugged in/faulty).
Power supply interrupted.
The displayed weight is per-
manently changing
Draught/air movement
Table/floor vibrations
Weighing plate has contact with foreign objects.
Electromagnetic fields / static charging (choose dif-
ferent location/switch off interfering device if possi-
ble)
The weighing result is obvi-
ously incorrect
The display of the balance is not at zero
Adjustment is no longer correct.
The balance is on an uneven surface.
Great fluctuations in temperature.
Warm-up time was ignored.
Electromagnetic fields / static charging (choose dif-
ferent location/switch off interfering device if possi-
ble)

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the KERN EWJ and is the answer not in the manual?

KERN EWJ Specifications

General IconGeneral
BrandKERN
ModelEWJ
CategoryScales
LanguageEnglish

Related product manuals