Page #25 background image
FCB-BA-e-1614
25
Dosing and
Zero-tracking
0.0
g
Vn
T
Tl
The Auto-Zero function is used to tare small variations in weight
automatically.
In the event that small quantities are removed or added to the
mater
ial to be weighed, incorrect weighing results can be
displayed due to the “stability compensation“ in the balance.
(Example: Slowly draining fluids from a container on the balance).
When apportioning involves small variations of weight, it is
advisable to switch off this function.
If Zero-Tracking
however is switched off, the weighing display
becomes more busy.
In weighing mode keep the PRINT key pressed until [Unit]
appears.
Press the MODE button several times until „tr“ is displayed.
Acknowledge using SET key, the current setting is displayed.
Select the desired settings by pressing the MODE key.
tr on
Function activated
tr off
Function deactivated
Use the SET key to confirm selection.
Selection
adjustment
weight
0.0
g
Vn
T
CAL
In the model series KERN FKB_A, the adjustment weight can be
selected from three/four pre-
set nominal values (approx.1/3; 2/3;
max) (refer also to following table, factory setting with grey
background). In order to achieve high-
quality weighing results in
the sense of the m
easuring technology, it is recommended to
select the nominal value as high as possible.
The non delivered
adjustment weights can be purchased from KERN as option.
In weighing mode keep the PRINT key pressed until [Unit]
appears.
Press the MODE key several times until „CAL“ is displayed.
Acknowledge using SET key, the current setting is displayed.
Select the desired settings by pressing the MODE key.
Use the SET key to confirm selection.

Other manuals for KERN FCB

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the KERN FCB and is the answer not in the manual?

KERN FCB Specifications

General IconGeneral
CalibrationExternal
Weighing plate materialStainless steel
Protection ClassIP54
Weighing unitsg, kg, lb, oz
DisplayLCD, with backlight
Power supplyAC Adapter / Batteries
Operating temperature10 °C - 40 °C
Minimum part weight at piece counting (Counting scale)0.1 g