22
Product Service: Domestic 1-800-522-7658, International 816-333-8811
Chapter 4: Using Your Freeze Dryer
Operation Checklist
The following checklist should be followed prior
to each use of your Freeze Dryer:
1. Wipe the interior of the collector chamber with
a soft cloth or paper towel to remove any
accumulated moisture.
2. Check the collector chamber drain hose to
insure that the hose is free of moisture and that
the drain plug is securely installed.
3. Using a soft, lint-free cloth or paper towel,
wipe the collector chamber lid gasket to
remove any dirt and contaminants that could
cause a vacuum leak. Vacuum grease is not
required on the lid gasket to obtain a proper
vacuum seal.
4. Remove the accessory drying chamber or
manifold from the connection port and using a
soft, lint-free cloth or paper towel, wipe the
port gasket and sealing surfaces of the dry
chamber/manifold to remove any dirt and
contaminants that could cause a vacuum leak.
Reinstall the drying chamber or manifold on
the port. Vacuum grease is not required on the
port gasket to obtain a proper vacuum seal.
5. Inspect each sample valve on the accessory
drying chamber or manifold and check for any
visible damage and for improper installation
that might cause a vacuum leak. Also check
that each sample valve is closed or in the
“vent” position.
Vacuum Pump Ballast
Setting
Most vacuum pumps are equipped with a gas
ballast mechanism. The freeze dry process
requires high vacuum (<133 x 10
-3
mBar or <100
microns). Therefore, it is recommended that the
gas ballast be closed during the operation of the
Freeze Dry System. If the gas ballast is left open
for extended periods of operation, the oil can be
pumped out the exhaust, causing the pump to fail.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Labconco FreeZone 6 and is the answer not in the manual?

Labconco FreeZone 6 Specifications

General IconGeneral
BrandLabconco
ModelFreeZone 6
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals