6
300680364_002_C0 - 12/2017 - © Leybold
Description
Interfaces
24 VDC / 1,5 A Power supply
Operate the device with functional extra low
voltage with positive isolation (PELV). Use
tested power supply.
RS485 TMP RS485 - Communication to the turbo pump
(start/stop, parameter read/write)
Max. cable length 30 m
USB TMP USB - Communication to the turbo pump
(start/stop, parameter read/write)
Max. cable length 5 m
ETH Connect a computer to the RJ-45 COM
interface and use the web server interface.
(Software updating, data logging etc.).
Max. cable length 30 m
Gauge 1 /
Gauge 2
At the TURBO.CONTROL i two vacuum
gauge heads can be connected. For this,
two of the following types can be used:
PTR 90, PTR 90N, TTR 91, TTR 91N,
TTR101, TTR 101N.
Max. cable length 30 m
1/3N Cell Buffer battery - 3V button cell (for real-time
clock and data logging).
Accessories
Mains adapter 800110V0027
Table housing 800110V0028
USB cable 2.0 type A/B, 1.8 m 800110V0108
RS485 cable 5 m 800103V0029
RS485 cable 1 m 800103V0027