- 12 -
LGE Internal Use OnlyCopyright © LG Electronics. Inc. All rights reserved.
Only for training and service purposes
4. Total Assembly line process
4.1. Adjustment Preparation
■ W/B Equipment condition
CA210 : CH14, Test signal : Inner pattern (80IRE) – in case of
LED back light
■ Above 5 minutes H/run in the inner pattern. (“power on” key of
adjust remote control)
► The spec of color temperature and coordinate
Model Mode Color Temp Color coordinate Remark
All
Cool (C50) 13,000k
x = 0.271 (±0.002)
y = 0.270 (±0.002)
※ Test signal
- Inner pattern for W/B adjust
- External white pattern
(80IRE, 204gray)
Medium (0) 9,300k
x = 0.286 (±0.002)
y = 0.289 (±0.002)
Warm (W50) 6,500k
x = 0.313 (±0.002)
y = 0.329 (±0.002)
► CA210 : CH 14, Test signal : Inner pattern (80IRE)
- Standard color coordinate and temperature using CA-1000 (by
H/R time)
▪ Normal line
model: (normal line)LN5xxx, LA6xxx, LA7xxx, LA8xxx
H/R Time(Min)
Cool Medium Warm
x y x x y x
271 270 286 289 313 329
1 0-2 283 287 298 306 322 342
2 3-5 282 285 297 304 321 340
3 6-9 281 284 296 303 320 339
4 10-19 279 281 294 300 318 336
5 20-35 277 277 292 296 316 332
6 36-49 275 274 290 293 314 329
7 50-79 273 272 288 291 312 327
8 80-119 272 271 287 290 311 326
9 Over 120 271 270 286 289 310 325
▪ Aging chamber line
model: (aging chamber)LN5xxx, LA6xxx, LA7xxx, LA8xxx
H/R Time(Min)
Cool Medium Warm
x y x x y x
271 270 286 289 313 329
1 0-5 282 285 297 304 321 340
2 6-10 278 280 293 299 317 335
3 11-20 275 275 290 294 314 330
4 21-30 272 272 287 291 311 327
5 31-40 269 269 284 288 308 324
6 41-50 268 267 283 286 307 322
7 51-80 267 266 282 285 306 321
8 81-119 266 264 281 283 305 319
9 Over 120 265 263 280 282 304 318
※ Connecting picture of the measuring instrument
(On Automatic control)
Inside PATTERN is used when W/B is controlled. Connect to
auto controller or push Adjustment R/C POWER-ON -> Enter
the mode of White-Balance, the pattern will come out.
● Auto-control interface and directions
(1) Adjust in the place where the influx of light like floodlight
around is blocked. (Illumination is less than 10ux).
(2) Adhere closely the Color Analyzer ( CA210 ) to the module
less than 10cm distance, keep it with the surface of the
Module and Color Analyzer’s Prove vertically.(80~100°).
(3) Aging time
- After aging start, keep the power on (no suspension of
power supply) and heat-run over 5 minutes.
- Using ‘no signal’ or ‘full white pattern’ or the others, check
the back light on.
● Auto adjustment Map(RS-232C)
RS-232C COMMAND
[ CMD ID DATA ]
Wb 00 00 White Balance Start
Wb 00 ff White Balance End
RS-232C COMMAND
[CMD ID DATA]
MIN
CENTER
(DEFAULT)
MAX
Cool Mid Warm Cool Mid Warm
R Gain jg Ja jd 00 172 192 192 192
G Gain jh Jb je 00 172 192 192 192
B Gain ji Jc jf 00 192 192 172 192
R Cut 64 64 64 128
G Cut 64 64 64 128
B Cut 64 64 64 128