5. Press button ‘O/I’ to return to standby mode. Wait for at least 5 minutes before
taking another measurement. If the device remains unused for 3 minutes, it will
return to standby mode automatically.
AUTOMATIC INFLATION
This device has 4 levels of inflation pressure: 190 mmHg, 230 mmHg, 270 mmHg
and 300 mmHg. When 190 mmHg is insufficient, or arm movement occurs, the
device will automatically inflate to an appropriate pressure level in order to obtain
the blood pressure measurement.
RAPID DEFLATION DURING MEASUREMENT
If you do not feel well during measurement, or want to stop measurement for any
reason, press button ‘O/I’. The cuff will quickly deflate and the device will switch to
standby mode.