WWW.MAXTEC.COM • (800) 748-5355 5 ENGLISH
1.0 INTRODUCTION
The MaxBlend 2 is a compact air/oxygen gas mixing device which incorporates the use of a
battery powered oxygen monitor. The gas mixing device (blender) provides precise mixing of
medical-grade air and oxygen, while the monitor measures the selected oxygen concentra-
tions from the blender’s gas flow and displays these measured concentrations on a digital
display. The monitor provides high and low alarm limits which, when exceeded, cause an
audible and visual alarm.
1.1 Indication for Use
The MaxBlend 2 is designed to provide a continuous air/oxygen gas mixture and to continu-
ously monitor the concentration of oxygen being delivered to infant, pediatric, and adult
patients. It is a restricted medical device intended for use by qualified, trained personnel,
under the direction of a physician, in professional healthcare settings, i.e., hospital, subacute,
and nursing-care facilities where the delivery and monitoring of air/oxygen mixtures is
required. This is not intended as a life supporting device.
1.2 MAX-550E Oxygen Sensor
The MAX-550E is a galvanic, partial pressure sensor that is specific to oxygen. It consists of two
electrodes (a cathode and an anode), a teflon membrane and an electrolyte. Oxygen diffuses
through the teflon membrane and immediately reacts electrochemically at a gold cathode.
Concurrently, oxidation occurs electrochemically at a lead anode, generating an electrical cur-
rent and providing a voltage output. Electrodes are immersed in a unique gelled weak acid
electrolyte, which is responsible for the sensor’s long life and motion insensitive characteristic.
Since the sensor is specific to oxygen, the current generated is proportional to the amount of
oxygen present in the sample gas. When no oxygen is present, there is no electrochemical
reaction, and therefore, negligible current is produced. In this sense, the sensor is self-zeroing.
CAUTION: The MAX-550E oxygen sensor is a sealed device containing a mild acid electro-
lyte and lead (Pb). These materials are hazardous waste constituents and should be disposed
of properly, or returned to Maxtec for proper disposal or recovery.
CAUTION: Dropping or severely jarring the sensor after calibration may shift the calibra-
tion point enough to require recalibration.
1.3 Symbol Guide
The following symbols and safety labels are found on the MaxBlend 2 and/or labeling:
ON/OFF Key Silent Key Backlight Key Unlock Key
Smart Alarm Key CAL (Calibration Key) Sleep Mode Indicator High Alarm Indicator
Low Alarm Indicator Calibration Reminder Smart Alarm Mode Indicator Alarm Silence Indicator
Below 18% Alarm Indicator Low Battery Indicator DOWN (Low Alarm Key) UP (High Alarm Key)
Do Not MR Unsafe Caution Warning
Corrosive Manufacturer
Federal law (USA) restricts
this device to sale by or
on order of a physician.
Type B Applied Parts
Attention, consult
accompanying documents
Ingress Protection Rating Liters Per Minute Oxygen Percent
Catalog Number Serial Number
Authorized Representative
in the European Community
Oxygen Enriched Air
Inlet Outlet Read Flow at Center of Ball Direct Current
Do not throw away. Follow
local guidelines for disposal
Meets ETL standards
50°C
(122°F)
-15°C
(5°F)
Storage Temperature Range Use No Oil
POWER SUPPLY SYMBOL GUIDE
The following symbols and safety labels are found on the MaxBlend 2 power supply (sold separately):
Two means of patient
protection (double insulated)
Conforms to EU requirements.
Combined UL / CSA Mark
For use in dry indoor
locations.
Combined UL / CSA Mark