8.0 SPECIFICATIONS
8.1 Base Unit Specifications
Measurement Range:....................................................0-100%
Resolution:.............................................................. 0.1%
Accuracy and Linearity: ..................1% of full scale at constant temperature, R.H. and
pressure when calibrated at full scale
Total Accuracy: .....................±3% actual oxygen level over full operating temp range
Response Time: ....................90% of final value in approximately 15 seconds at 23˚C
Warm-up Time: ................................................... None required
Operating Temperature: .....................................15˚C - 40˚C (59˚F - 104˚F)
Storage Temperature:.......................................-15˚C - 50˚C (5˚F - 122˚F)
Atmospheric Pressure:............................................ 800-1013 mBars
Humidity:................................................ 0-95% (non-condensing)
Power Requirements:..............................2, AA Alkaline batteries (2 x 1.5 Volts)
Battery Life: ..............................approximately 5000 hours with continuous use
Low Battery Indication:...................................."BAT" icon displayed on LCD
Sensor Type: ................................. Maxtec MAX-250 series galvanic fuel cell
Expected Sensor Life:........................... > 1,500,000 O
2
percent hours minimum
(2-year in typical medical applications)
Dimensions:
A Model Dimensions: ................3.0”(W) x 4.0”(H) x 1.5”(D) [76mm x 102mm x 38mm]
A Weight:........................................................0.4 lbs. (170g)
AE Model Dimensions: ..............3.0”(W) x 36.0”(H) x 1.5”(D) [76mm x 914mm x38mm]
Height includes external cable length (retracted)
AE Weight:.......................................................0.6 lbs. (285g)
Drift of Measurement: .......................< +/-1% of full scale at constant temperature,
pressure and humidity
8.2 Sensor Specifications
Type: .............................................. Galvanic fuel sensor (0-100%)
Life: ................................................2-years in typical applications
9.0 MAXO
2
+ SPARE PARTS AND ACCESSORIES
9.1 Included With Your Unit
Part Number Item
R217M72 User’s Guide and Operating Instructions
RP76P06 Lanyard
R110P10-001 Flow Diverter
RP16P02 Blue Tee Adapter
R217P35 Dovetail Bracket

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Maxtec MaxO2 Plus and is the answer not in the manual?

Maxtec MaxO2 Plus Specifications

General IconGeneral
BrandMaxtec
ModelMaxO2 Plus
CategoryAnalytical Instruments
LanguageEnglish

Related product manuals