A - 6
360 J defibrillation waveform into impedance of 25Ω, 50Ω, 75Ω, 100Ω, 125Ω, 150Ω, 175Ω
Selected energy accuracy
Impedance
Energy
25Ω 50Ω 75Ω 100Ω 125Ω 150Ω 175Ω Accuracy
10 J 9.7 J 10 J 9.7 J 9.3 J 8.9 J 8.5 J 8.1 J ±10% or ±2J,
whichever is
greater
15 J 15 J 15 J 15 J 14 J 13 J 13 J 12 J
20 J 20 J 20 J 20 J 19 J 18 J 17 J 16 J
30 J 29 J 30 J 29 J 28 J 27 J 25 J 24 J
50 J 49 J 50 J 49 J 47 J 45 J 43 J 41 J
70 J 68 J 70 J 68 J 65 J 62 J 60 J 57 J
100 J 97 J 100 J 97 J 93 J 89 J 85 J 81 J
150 J 146 J 150 J 146 J 140 J 134 J 128 J 122 J
170 J 166 J 170 J 166 J 159 J 151 J 145 J 138 J
200 J 195 J 200 J 195 J 187 J 178 J 170 J 163 J
300 J 292 J 300 J 292 J 280 J 267 J 255 J 244 J
360 J 351 J 360 J 350 J 336 J 321 J 306 J 293 J
Charge time (at 20 °C± 5 °C of ambient temperature)
Battery status From lid opened to
charge done
From initiation of rhythm
analysis to charge done
From initial power
on to charge done
200J 360J 200J 360J 200J
New battery <8 s <15 s <5 s <12 s <7 s
New battery after 15 times of
360J discharges
<8 s <15 s <5 s <12 s <7 s
Time (ms)
Voltage (V)

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray BeneHeart S1 and is the answer not in the manual?

Mindray BeneHeart S1 Specifications

General IconGeneral
BrandMindray
ModelBeneHeart S1
CategoryMedical Equipment
LanguageEnglish

Related product manuals