Acoustic Output Reporting Table(61157-2007)
F-5
F.5 65EC10EB
Mode
Parameter
B M B+M
System settings
Standard used: IEC 61157 Ed2
Display depth2.16cm
focus 2cm AP100%
Display depth2.16cm
focus 2cm AP100%
Display depth2.16cm
focus 2cm AP100%
p
r
(MPa)
2.18±0.36 2.18±0.36 2.18±0.36
I
spta
(mW/cm
2
)
2.43±0.80 57.21±18.87 59.61±19.66
I
ob
(mW/cm
2
)
2.76±0.91 0.40±0.13 3.93±1.30
Power output (mW)
9.07±0.49 1.33±0.07 12.93±0.71
Output beam dimensions
a
(׎)
(mm)
20.46 20.46 20.46
Z
p
(mm)
1.44±0.25 1.44±0.25 1.44±0.25
w
12
(| |) (mm)
2.65±0.13 3.64±0.13 2.65±0.13
w
12
() (mm)
1.24±0.13 2.48±0.13 1.24±0.13
f
awf
(MHz)
6.59±0.17 6.59±0.17 6.59±0.17
prr (KHz)
- 1,000.00±1.00 1,000.00±1.00
srr (Hz)
45.00±1.00 - 16.00±1.00
Z
tt
(mm)
1.19 1.19 1.19
Z
ts
(mm)
contact contact contact
Acoustic output freeze
Yes Yes Yes
Inclusive modes
- - -
a: Deq ,equivalent aperture diameter,in IEC60601-2-37