© National Instruments Corporation 13 NI 446X Calibration Procedure
4. Modify the AI voltage channel property using the following functions.
5. Configure the timing properties of the acquisition using the following
function.
6. Start the task using the following function.
NI-DAQmx Function Call LabVIEW Block Diagram
Call DAQmxSetAIGain with the
following parameters:
taskHandle:
myTaskHandle
channel: dev1/ai0
data: -20.0
Call DAQmxSetAICoupling with the
following parameters:
taskHandle:
myTaskHandle
channel: dev1/ai0
data: DAQmx_Val_DC
NI-DAQmx Function Call LabVIEW Block Diagram
Call DAQmxCfgSampClkTiming with
the following parameters:
taskHandle:
myTaskHandle
source: NULL
rate: 204800.0
activeEdge: DAQmx_Val_Rising
sampleMode:
DAQmx_Val_FiniteSamps
sampsPerChanToAcquire: 102400
NI-DAQmx Function Call LabVIEW Block Diagram
Call DAQmxStartTask with the following
parameters:
taskHandle:
myTaskHandle

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the National Instruments 446 Series and is the answer not in the manual?

National Instruments 446 Series Specifications

General IconGeneral
BrandNational Instruments
Model446 Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals