b) Input the NI 4113 measurements as the reported outputs.
c) Input the specified level range as the range.
9. Update the output calibration constants by configuring and calling the niDCPower Cal
Adjust Current Limit VI or niDCPower_CalAdjustCurrentLimit function.
a) Input the calculated shunt current measurements as the measured outputs.
b) Input the test points as the requested outputs.
c) Input the specified level range as the range.
10. If more than one level range is specified, repeat the previous steps using the values
specified in each level range.
Related Information
For information about VIs or functions, refer to the programming reference of the NI DC
Power Supplies and SMUs Help.
Closing the Adjustment Session
Close the session and commit the new constants to hardware by calling the niDCPower
Close External Calibration VI or niDCPower_CloseExtCal function and specifying
Commit as the calibration close action.
Reverification
Repeat the Verification section to determine the as-left status of the device.
Note If any test fails reverification after performing an adjustment, verify that you
have met the Test Conditions before returning your device to NI. Refer to the
Worldwide Support and Services section for information about support resources or
service requests.
Related Information
Test Conditions on page 3
Worldwide Support and Services
The NI website is your complete resource for technical support. At ni.com/support, you have
access to everything from troubleshooting and application development self-help resources to
email and phone assistance from NI Application Engineers.
Visit ni.com/services for NI Factory Installation Services, repairs, extended warranty, and
other services.
Visit ni.com/register to register your NI product. Product registration facilitates technical
support and ensures that you receive important information updates from NI.
A Declaration of Conformity (DoC) is our claim of compliance with the Council of the
European Communities using the manufacturers declaration of conformity. This system
affords the user protection for electromagnetic compatibility (EMC) and product safety. You
NI PXIe-4113 Calibration Procedure | © National Instruments | 17

Other manuals for National Instruments NI PXIe-4112

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the National Instruments NI PXIe-4112 and is the answer not in the manual?

National Instruments NI PXIe-4112 Specifications

General IconGeneral
BrandNational Instruments
ModelNI PXIe-4112
CategoryPower Supply
LanguageEnglish

Related product manuals