NI PXI-4110 Calibration Procedure | © National Instruments | 11
To ensure the system has had adequate time to settle, wait three seconds after requesting a new
current before taking a measurement with the DMM. The current programming test limits in
Table 7 are derived from the current programming accuracy specifications.
Table 7. NI 4110 Output Parameters and Test Limits for Current
Programming Accuracy Verification
Channel(s)
Output
(V)
Current
Range
Iteration
Equivalent
Resistance
()
Output
(A)
Test Limit
(A)
0 6 1A 1 150 0.020000 ±0.004030
2 12 0.250000 ±0.004375
3 6 0.500000 ±0.004750
4 4 0.750000 ±0.006063
5 3 1.000000 ±0.008000
1 16 1A 1 500 0.020000 ±0.004030
2 40 0.250000 ±0.004375
3 20 0.500000 ±0.004750
4 13 0.750000 ±0.006063
5 9 1.000000 ±0.008000
1 16 20 mA 1
25,000 0.000400 ±0.000060
2 2,000 0.005000 ±0.000068
3 1,000 0.010000 ±0.000075
4 670 0.015000 ±0.000083
5 500 0.020000 ±0.000090
2 -16 1A 1 500 0.020000 ±0.004030
2 40 0.250000 ±0.004375
3 20 0.500000 ±0.004750
4 13 0.750000 ±0.006063
5 9 1.000000 ±0.008000

Other manuals for National Instruments PXI-4110

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the National Instruments PXI-4110 and is the answer not in the manual?

National Instruments PXI-4110 Specifications

General IconGeneral
BrandNational Instruments
ModelPXI-4110
CategoryPower Supply
LanguageEnglish

Related product manuals