NI PXI-4110 Calibration Procedure | © National Instruments | 5
Use shielded copper wire for all cable connections to the device. Use twisted-pair wire to
eliminate noise and thermal offsets.
Plug the chassis and the instrument standard into the same power strip to avoid ground
loops.
Calibration Procedures
The calibration process includes the following steps:
1. Initial Setup—Install the device and configure it in Measurement & Automation Explorer
(MAX).
2. VerificationVerify the existing operation of the device. This step confirms whether the
device is operating within its specified range prior to calibration.
3. Adjustment—Perform an external adjustment of the device that adjusts the calibration
constants with respect to a known voltage source. The adjustment procedure automatically
stores the calibration date on the EEPROM to allow traceability.
4. ReverificationRepeat the verification procedure to ensure that the device is operating
within its specifications after adjustment.
These steps are described in more detail in the following sections.
Note In some cases, a complete calibration procedure may not be required. Refer
to Where to Go for Support for more information.
Initial Setup
Refer to the NI PXI-4110 Getting Started Guide for information about how to install the
software and hardware and how to configure the device in MAX.
Tables 3 and 4 list configuration information for the calibration equipment required for
verification.
Table 3. Calibration Equipment Configuration for Voltage Programming and Measurement
Verification/Adjustment
NI 4110 Channel(s)
DMM*
Function Range
Input Impedance
0
DC Voltage
10 V 10 G
1, 2 100 V 10 M
* Use the highest resolution available on the DMM.
Assumes an NI 4071 DMM. For all other DMMs, use the range and input impedance closest to the
values listed in this table.

Other manuals for National Instruments PXI-4110

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the National Instruments PXI-4110 and is the answer not in the manual?

National Instruments PXI-4110 Specifications

General IconGeneral
BrandNational Instruments
ModelPXI-4110
CategoryPower Supply
LanguageEnglish

Related product manuals