14 | ni.com | NI PXI-4110 Calibration Procedure
Verifying Load Regulation
Complete the load regulation test to verify that the output voltage falls within specified limits
when the load current changes or to verify that the output current falls within specified limits
when the load voltage changes. Each test requires two resistors to vary the load voltage or
current.
Note Load Voltage Verification results are dependent upon voltage and current
measurement accuracy. Any failure of the Voltage Measurement Verification or
Current Measurement Verification steps may result in a failure of the Load
Regulation Verification steps.
Connect the NI APS-4100 to the Aux Power Input connections for load regulation tests.
1, 2 20 mA 1 open 0
±0.000012 ±0.000035
2 1,000 ±5 ±0.000020 ±0.000043
3 1,000 ±10 ±0.000027 ±0.000050
4 1,000 ±15 ±0.000035 ±0.000058
5 1,000 ±20 ±0.000042 ±0.000065
*
When verifying channel 1, the Output is positive. When verifying channel 2, the Output is negative.
To facilitate an accurate measurement using an Output value of 0, verify that the DMM and the external
load are disconnected from the NI 4110. For all other Output values, the DMM and the external load are
connected to the NI 4110.
Table 8. 4110 Output Parameters and Test Limits for Current Measurement
Accuracy Verification (Continued)
Channel(s)
Current
Range
Iteration
Equivalent
Resistance
()
Output
(V)
*
Test Limit
(A)
Published
Specification
(A)

Other manuals for National Instruments PXI-4110

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the National Instruments PXI-4110 and is the answer not in the manual?

National Instruments PXI-4110 Specifications

General IconGeneral
BrandNational Instruments
ModelPXI-4110
CategoryPower Supply
LanguageEnglish

Related product manuals