19 Annex
52 / 60 Instruction Manual Easyfield® (GA/15000/XXXX/EN 1219 Rev01)
To be in compliance with the requirements of the IEC 60601-1-2. 6.1 and
6.2 the following types of equipment, accessories, power adapters and
cables must be used.
19.2 Guidance and Manufacturer’s Declaration -
Electromagnetic Emmisssions and Immunity for the Easyfield®
Order number Description
15000 OCULUS-Easyfield® C
15005 OCULUS-Easyfield® S
05200320 Cable with connector plug,
EU standard
2.5m
05200210 ( 110 Volt) Cable with connector plug,
US standard
2.5m
015692000010 USB FS Med isolator
05150805 Power supply
GSM40B12-P1J
see “Power
adapter”
page 49
Guidance and manufacturer´s declaration electromagentic emmissions
IEC 60601-1-2, 5.2.2.1, table 1
The OCULUS Easyfield® is intended for operation in the electromagnetic environment specified below. The user
of the Easyfield® should ensure that it is being used in such an environment.
Emissions test Compliance Electromagnetic environment - guidance
RF emissions
CISPR 11
Group 1 The Easyfield® uses RF energy only for its internal func-
tion. Therefore, its RF emissions are very low and are not
likely to cause any interference in nearby electronic
equipment.
HF-emissions
CISPR 11
Class B
Harmonics emissions
IEC 61000-3-2
Class A
Voltage fluctuations /
flicker emissions
IEC 61000-3-3
complies

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the OCULUS Easyfield C and is the answer not in the manual?

OCULUS Easyfield C Specifications

General IconGeneral
BrandOCULUS
ModelEasyfield C
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals