19 Annex
Instruction Manual Easyfield® (GA/15000/XXXX/EN 1219 Rev01) 55 / 60
Recommended separation distances between portable and mobile
RF communications equipment and the Easyfield®,
IEC 60601-1-2, 5.2.2.2, table 6
The Easyfield® in intended for us e in an electromagnetic environment in which radiated RF disturbances are
controlled. The customer or the user of the Easyfield® can help prevent electromagnetic interference by main-
taining a minimum distance between portable and mobile RF communications equipment (transmitters ) and
the Easyfield® as recommended below, according to the maximum output power of the communications equip-
ment.
Rated maximum output pow-
er of transmitter
W
Separation distance according to frequency of transmitter m
150 KHz to 80 Mhz
d= 1.2 P
80 MHz to 800 MHz
d= 1.2 P
800 MHz to 2.5 GHz
d= 2.3 P
0.01 0.12 0.12 0.23
0.1 0.38 0.38 0.73
1 1.2 1.2 2.3
10 3.80 3.80 7.3
100 12 12 23
For transmitters rated at a maximum output power not listed above, the recommended separation distance d
in metres (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is
the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption
and reflection from structures, objects and people.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the OCULUS Easyfield C and is the answer not in the manual?

OCULUS Easyfield C Specifications

General IconGeneral
BrandOCULUS
ModelEasyfield C
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals