FLUSHING PUMP OFP
10
5 CARE AND STO RAGE
5.1 Care
(1) Clean the OFP with a soft cloth or gauze moistened with a neutral pH detergent, rinse with clean water, then wipe
over with a cloth dampened with 70% alcohol.
CAUTION
Do not use harsh or abrasive cleaning materials on the OFP as damage may result.
Ensure all surfaces are thoroughly dried before reusing the OFP.
Ensure footswitch is removed from the floor and stored safely to prevent it being damaged.
(2) Inspect the fluid c ontainer and fluid container cap before each session of use for obvious signs of damage, replace
if necessary.
(3) The fluid container and fluid container cap may be washed with a neutral pH detergent, then thoroughly rinsed
prior to autoclaving.
(4) Reprocess the MAJ--855 following each patient use accor ding to the manufacturerās instructions supplied with the
product.
(5) The fluid container and fluid container cap should be autoclaved at the end of the day (see Section 4).
5.2 Storage
(1) Remove the irr igation tubing and filter from the peristaltic pump and discard. Disconnect the power lead. Empty the
fluid container and autoclave prior to storage.
(2) The OFP should be stored within the environmental conditions given in Section 8.
6 MAINTENANCE AND REPAIR
6.1 Routine Maintenance
The following routine maintenance should be performed at the intervals specified:
User - Daily:
At the beginning of each day of use, check that the pump head safety switch is operating correctly as follows:
With no tube fitted, operate the pump at maximum speed and, taking care to keep fingers clear of the rollers, lift the
pump head cover. The rollers should promptly stop rotating. If they do not, s witch off the pump, disconnect it from the
electrical power supply and contact your Olympus representative.
User - Monthly:
Check the pump head and r ollers for damage.
Check that the power cable is in good condition.
Hospital Engineer or Olympus - Annually:
WARNING
Lethal electrical hazard -- ensure the unit is disconnected from the power supply before attempting to enter the
unit for inspection. This routine maintenance procedure must only be performed by a suitably qualified
technician or authorized Olympus service personnel.
Check that the power cable is in good condition.
Inspect the unit externally and perform safety checks as detailed in the maintenance & repair manual.