ERROR INDICATORS
ERROR INDICATORS AND TROUBLESHOOTING TIPS
SYMBOL CAUSE CORRECTION
Air plug disconnected.
Insert the plug securely. Refer to “Applying the
Arm Cuff”.
Arm cuff not applied correctly.
Apply the arm cuff correctly. Refer to “Applying
the Arm Cuff”.
Air is leaking from the arm cuff. Replace the cuff with the new one.
Cuff over or under inflated.
Repeat measurement. Remain still and do not
talk during measurement. Refer to “Taking a
Measurement”.
If “E2” appears repeatedly, inflate the cuff manually
until it is 30 to 40 mmHg above your previous
measurement result. Refer to “Taking a Measurement”.
The arm cuff was inflated above 299
mmHg when inflating the cuff manually.
Do not inflate the cuff above 299 mmHg. Refer to
“Taking a Measurement”.
Movement during measurement.
Repeat measurement. Remain still and do not
talk during measurement. Refer to “Taking a
Measurement”.
Clothing is interfering with the
arm cuff.
Remove any clothing interfering with the arm cuff.
Refer to “Applying the Arm Cuff”.
Device error. Contact Customer Service.
Cuff is not applied correctly. Refer to “Applying the Arm Cuff”.
Batteries are worn.
Replace the four batteries.
Refer to “Battery Installation”.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Omron BP760 and is the answer not in the manual?

Omron BP760 Specifications

General IconGeneral
TypeAutomatic
Cuff size420 cm
Display typeLCD
Date imprintingYes
Number of users1 user(s)
Placement supportedUpper arm
Mean arterial pressure-
Battery typeAA
Battery voltage1.5 V
Number of batteries supported4
Weight and Dimensions IconWeight and Dimensions
Depth86 mm
Width158 mm
Height124 mm