EasyManuals Logo
Home>Omron>Blood Pressure Monitor>M2

Omron M2 User Manual

Omron M2
35 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #20 background imageLoading...
Page #20 background image
20
3.Using the Unit
3.4
Instructions for Special Conditions
If your systolic pressure is known to be more than 220 mmHg, press
and hold the O/I START button until the arm cuff inflates 30 to 40 mmHg
higher than your suspected systolic pressure.
1. Press the O/I START button
to turn the unit on.
Measurement starts.
2. When the cuff starts to inflate, press the O/I START button
and keep it pressed until the pressure is 30 to 40 mmHg
higher than your suspected systolic pressure.
Note: You cannot inflate the cuff above 299 mmHg. (An error
will be displayed if you try to inflate the cuff above
299 mmHg.)
3. Release the O/I START button when the cuff has been inflated
to the desired pressure.
The cuff starts to deflate and measurement starts.
4. The rest of the procedure is the same as for normal
measurement. Refer to 3.3, steps 2 to 6.
Note: Do not apply more pressure than necessary.
HEM-7117-E_main.book Page 20 Friday, May 14, 2010 8:43 AM

Other manuals for Omron M2

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Omron M2 and is the answer not in the manual?

Omron M2 Specifications

General IconGeneral
TypeAutomatic
Cuff size22 - 32 cm
Display typeLCD
Placement supportedUpper arm
Pulse rate measuring-
Mean arterial pressureYes

Related product manuals