28
,QFDVHRIDQ\RIWKHEHORZSUREOHPVRFFXUGXULQJXVH¿UVWFKHFNWKDWQRRWKHUHOHFWULFDOGHYLFHLV
within 12 inches (30cm).
If the problem persists, please refer to the table below.
Problem Possible cause Solution
The nebulization rate is
extremely low.
Batteries are low (the orange
EDWWHU\ORZLQGLFDWRUÀDVKHV
Replace the batteries with new
ones. Refer to sub-section 2.3.
The mesh is stained or clogged. 1. Clean the mesh. Refer to sub-
section 2.2.
2. Replace the mesh cap with a new
one. Refer to sub-section 2.5.
The green power
indicator does not turn
on and the unit does not
nebulize.
The batteries are depleted (the
orange battery low indicator turns
on).
Replace the batteries with new ones
or use the AC adapter(Optional).
Refer to sub-section 2.3 and 2.4.
The capacity of rechargeable
batteries is low.
Charge the batteries with a
commercially available recharger.
The batteries are inserted in the
wrong direction.
Re-insert the batteries in the correct
direction. Refer to sub-section 2.3.
The AC adapter (Optional) is not
connected correctly to a power
outlet.
Check that the AC adapter is
connected to a power outlet and
main unit. Unplug then reinsert the
plug if necessary.
The contact on the AC adapter
(Optional) is contaminated.
:LSHDZD\DQ\FRQWDPLQDWLRQZLWKD
cloth that has been lightly moistened
with water or sanitizing alcohol. Dry
completely before use.
4. Troubleshooting