10
2. Check the displayed waveforms and regulate the probe till a correct
compensation is achieved (see Figure 7-2 and Figure 7-3).
Overcompensated Compensated correctly Under compensated
Figure 7-2 Displayed Waveforms of the Probe Compensation
3. Repeat the steps mentioned if needed.
Figure 7-3 Adjust Probe
8. How to Use the Probe Safely
The safety guard ring around the probe body protects your finger against any
electric shock, shown as Figure 8-1.
Figure 8-1 Finger Guard
Warning:
To avoid electric shock, always keep your finger behind the safety
guard ring of the probe during the operation.
To protect you from suffering from the electric shock, do not touch any
metal part of the probe tip when it is connected to the power supply.
Before making any measurements, always connect the probe to the
instrument and connect the ground terminal to the earth.