EasyManuals Logo
Home>PerkinElmer>Laboratory Equipment>Lambda 950

PerkinElmer Lambda 950 User Manual

PerkinElmer Lambda 950
82 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #31 background imageLoading...
Page #31 background image
System Description . 31
Mirrors 14 and 15 are used to select the appropriate detector. The mirror translation unit
(Figure 11) has three positions. At position 1, mirror M14 (the aluminum, SiO
2
-coated mirror)
directs the beam to the PMT detector at the base of the compartment. At position 2, both
mirrors are outside of the beam path, and the beam goes to the InGaAs detector. At position
3, mirror M15 (the gold mirror) directs the beam to the PbS detector at the top of the optics
compartment. The changeover between the detectors occurs at predefined wavelengths, and
is controlled by the software.
Figure 11 The arrangement of the detectors of the Three Detector Module
Sphere-only Systems
In addition to the standard configurations described above, the Lambda 650/750/950/1050
spectrometers can be supplied with an integrating sphere only, containing a PMT detector for
the UV-vis range and either a PbS or InGaAs detector for the NIR range. These systems have
simpler optical designs than the standard configurations, and therefore the spectral scanning
ranges in the UV/Vis and NIR regions are slightly reduced. The detector options and range
specifications depend upon the spectrometer model; consult your PerkinElmer Service
Representative for more details. Further information on the sphere detectors is available in
the
Sphere Accessories User’s Guide
(L6050011); we recommend that you read this user’s
guide before installing your sphere accessory and switching on the power to the
spectrometer.
Detector-less Systems
All of the high performance Lambda spectrometer systems can also be supplied with no
detectors included, although a detector must be installed before the system can be operated.
Consult your PerkinElmer Service Representative for more details.
Mirror 15
PMT detector
Mirror 11
Mirror 12
InGaAs
detector
PbS detector
Mirror translation
unit
Mirror 14
Mirror 12

Table of Contents

Other manuals for PerkinElmer Lambda 950

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the PerkinElmer Lambda 950 and is the answer not in the manual?

PerkinElmer Lambda 950 Specifications

General IconGeneral
BrandPerkinElmer
ModelLambda 950
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals