English
14
CE Certification
This instrument conforms to the following standards:
EN50081-1:1992, Electromagnetic Emissions
EN50082-1:1997, Electromagnetic Susceptibility
Tests were conducted using a frequency range of 80-1000 MHz with the instrument in
three orientations. The average error for the three orientations is ±1.4°C (±2.5°F) at 3
v/m throughout the spectrum. However, between 162 MHz and 792 MHz at 3 V/m, the
instrument may not meet its stated accuracy.
Troubleshooting
Code Problem Action
– – – (on display) Target temperature Select target within specifications
is over or under range
Battery icon
appears Low battery Check and/or replace battery
Blank display Possible dead battery Check and/or replace battery
Laser doesn’t work (1) Low or dead battery (1) Replace battery
(2) Ambient temperature (2) Use in area with lower ambient
above 40°C (104°F) temperature
ERR Possible damage Contact your distributor
by EMF
1.888.610.7664 info@Raytek-Direct.com

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the RayTek Raynger ST60 ProPlus Standard and is the answer not in the manual?

RayTek Raynger ST60 ProPlus Standard Specifications

General IconGeneral
BrandRayTek
ModelRaynger ST60 ProPlus Standard
CategoryThermometer
LanguageEnglish

Related product manuals