Specifications
Temperature range D : S*
ST60 ProPlus™ -32 to 600°C (-25 to 1100°F) 30 : 1
ST80 ProPlus™ -32 to 760°C (-25 to 1400°F) 50 : 1
Contact Probe -40 to 260°C (-40 to 500°F)
Display Resolution 0.1°C (0.1°F)
Accuracy For targets:
(infrared -- assumes Above 23°C (73°F): ± 1% of reading or ± 1°C (± 2°F),
ambient operating whichever is greater
temp. of 23-25°C[73-77°F]) -18 to 23°C (0 to 73°F): ± 2°C (± 3°F)
@ calibration geometry** -26 to -18°C (-15 to 0°F): ± 2.5°C (± 4°F)
-32 to -26°C (-25 to -15°F): ± 3°C (± 5°F)
Probe ≤ ± 1°C (± 2°F) or 1% of reading, whichever is greater
Repeatability ± 1°C (± 2°F), or ± 0.5% of reading, whichever is greater
Response time 500 mSec
Spectral response 8–14 µm
Emissivity Digitally adjustable from 0.10 to 1.0
Ambient operating range 0 to 50°C (32 to 120°F)
Laser Max Laser turns off above 40°C (104°F) ambient temperature
Relative humidity 10–95% RH noncondensing, at < 30°C (86°F) ambient
Storage temperature -20 to 60°C (-13 to 158°F) without battery
Weight / Dimensions 320g (11oz); 200 x 160 x 55 mm (8 X 6 X 2 in)
Power 9V Alkaline or NiCd battery
Typical Battery life 20 hours with laser and backlight on 50%
(Alkaline) 40 hours laser and backlight off
Tripod Mount 1/4” 20 UNC threading
*90% encircled energy at the focal point. (Focal point = 914mm (36 inches)).
**Unit is calibrated at 279.4mm (11 in) away from 139.7mm (5.5 in) diameter, 0.95 emissivity blackbody.
Some Noncontact Thermometer Uses
Electrical/Industrial •
Industrieanwendungen • Électricité-Méchanique
•Uso electromecánico Industrial • Manutenção Elétrica e Industrial
••
Automotive & Diesel •
KFZ-Instandesetzung und Wartung
• Vehicules (essence et diesel) • Automotriz y Diesel • Automotivo e Diesel
••
Food Safety •
Lebensmittelüberwachung
•
Hygiène alimentaire
• Seguridad en Alimentos • Segurança de Alimentos Pereciveis
••