Chapter 6 MATH and Measurement RIGOL
MSO1000Z/DS1000Z User’s Guide 6-11
"AND" Operation
Perform logic "AND" operation on the waveform voltage values of the specified
sources point by point and display the results. In operation, when the voltage value
of the source channel is greater than the threshold of the corresponding channel, it is
regarded as logic "1"; otherwise logic "0".
The results of logic AND operation of two binary bits are as follows.
Table 6-2 Logic "AND" Operation
A B A&&B
0 0 0
0
1
0
1
0
0
1
1
1
Press MATH Math Operator to select "A&&B":
Press Operation to enable or disable the "AND" operation function.
Press SourceA and SourceB to select the desired channels (CH1-CH4 or
D0-D15).
If source A selects any channel in CH1-CH4, press Thre.A and use set
the threshold of source A in logic operation. If source A selects any channel
in D0-D15, Thre.A will be hidden automatically.
If source B selects any channel in CH1-CH4, press Thre.B and use set
the threshold of source B in logic operation. If source B selects any channel
in D0-D15, Thre.B will be hidden automatically.
Press Offset and use to adjust the vertical position of the operation results.
Press Scale and use to adjust the vertical scale of the operation results.
Press Scale Reset to adjust the vertical scale of the operation results to the
optimal value according to the current configuration.
Press Options to set the start and end points of the operation results, enable or
disable waveform invert, etc.
Press Start and use to set the start point of the operation results.
Press End and use to set the end point of the operation results.
Press Invert to enable or disable the inverted display function of the
waveform.
Press Sens. and use to set the sensitivity of the digital signal converted
from the analog signal on the source. The setting range is from 0 Div to
0.96 Div.
Press Auto Scale to enable or disable the auto scale function. When auto
scale is enabled, the instrument will adjust the vertical scale of the
operation results to the optimal value according to the current
configuration.
www.GlobalTestSupply.com
Find Quality Products Online at: sales@GlobalTestSupply.com

Table of Contents

Other manuals for Rigol MSO1104Z

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Rigol MSO1104Z and is the answer not in the manual?

Rigol MSO1104Z Specifications

General IconGeneral
BrandRigol
ModelMSO1104Z
CategoryTest Equipment
LanguageEnglish

Related product manuals