RIGOL Chapter 1 Disassembly and Assembly
1-14 MSO5000-E Service Guide
Disassemble the Power Key
Figure 1-11 Disassemble the Power Key
Part Description:
2 screws (M3x8 torx pan head combination screw, used for fastening the power switch
component)
Disassembly Steps:
1. Use the T10 screwdriver to remove 2 screws () that secure the power switch component (power
key + mechanical switch).
2. Remove the power switch component from the metal front panel.
3. Remove the power key from the mechanical switch.
Power Key
Mechanical Switch