Maintenance cobas c pack opener
a Changing the fuses
1
Switch off the unit and disconnect the power cord.
If the power cord is still connected, you will not be able to access the fuse holder.
2
Using the tip of a blade screwdriver, lift the fuse holder out of the unit.
3
Replace both fuses.
Make sure you are using the correct types:
(Type 2A / 250V AC (5 x 20 mm) - Blow Characteristic -Time Delay).
4
Reconnect the power cord to the unit.
5
Switch on the unit.
Figure 8 Lifting the fuse holder