Operator’s Manual · Version 1.2 17
cobas c pack opener Maintenance
Maintenance
The cobas c pack opener is a maintenance-free device.
Users are responsible for cleaning the unit on a regular basis.
a Empty the waste container and clean the unit
1
Tilt the waste container and pull out.
2
Dispose of the caps in accordance with local specifications and laws.
3
Put the waste container in the dishwasher.
4
Clean surfaces of the unit with a lint-free tissue or a moistened cotton bud if
necessary, using mild detergents or disinfectants, preferably 70% alcohol.
NOTICE
Always switch off and unplug the unit prior to opening the waste container.
Always switch off and unplug the unit prior to cleaning.
Do not allow liquids to enter the device.
Do not spray cleaning agents directly onto the device. Use a lint-free tissue or a moistened
cotton bud.
Figure 7 Emptying the waste container
After max. 150 cobas c packs large, empty the waste container.
The waste container is dishwasher-safe up to 65°C (149°F).
Do not put adapters into the dishwasher!