WARNING: A 20% (20 mL/min) level of irrigation is
required at a minimum for the handpiece. Lower levels
may result in inconsistent irrigation that may lead to
heat generation in the handpiece and at the surgical
site, increasing the risk of clogging and abnormal wear
of the accessory. Refer to the instructions for use
supplied with the CORE 2 system for the procedure on
how to adjust the irrigation level.
8. Using the footswitch, activate the handpiece. Depending on
user preference, the footswitch pedal may be assigned different
functions using the console.
WARNING: DO NOT resect tissue with the Pello
Microdebrider system while actively tracking with
navigation systems. The endoscopic monitor shall be
used while dissecting.
9. The jog function provided by the console allows the accessory
to rotate slowly and may be used to align the cutting window to
the desired position.
WARNING: DO NOT operate the handpiece in the jog
function mode for more than one minute. Failure to
comply may result in overheating.
CAUTION: DO NOT operate the handpiece if the accessory does
not rotate or oscillate when power is applied to the handpiece.
Failure to comply may damage the handpiece.
Clearing Obstructions
Remove the accessory from the handpiece and gently insert
the ENT brush stylet (provided in the packaging) into the cutting
window or the proximal opening and guide along the inner lumen of
the accessory. Use light pressure to remove debris.
NOTE: Ensure the cutting tip is in the fully open position before
inserting the stylet.
Final Disassembly
1. Remove the accessory. See the To Remove an Accessory
section.
2. Unplug the handpiece cord from the console.
3. Disconnect the suction and irrigation tubes from the handpiece.