14 www.stryker.com
EN 700000890732 Rev-AE
Specifications
WARNING: ALWAYS check any documentation that
accompanies attachments and/or accessories for
special duty cycle and usage instructions.
Model: Pello Microdebrider
REF 6290-100-000
Dimensions: 7.1 inch [18.2 cm] length (handpiece only)
1.1 inch [2.8 cm] height
0.9 inch [2.3 cm] width
Mass: 0.84 lb [380 g] (including cord)
Speed: Oscillation mode – 1400 to 5000 rpm
Forward/reverse mode – 1400 to 12000
rpm
Mode of Operation:
Duty Cycle:
Rest Between Cycles:
Non-continuous Operation
3 minutes on/3 minutes off, 10 times
3 hours
Power Supply: Stryker CORE 2 Console
40 V
Equipment Type:
Type BF applied part
Applied Part: The distal end of the handpiece as
defined by the manufacturer.
Maximum
Temperature of
Applied Part:
Less than or equal to 48°C [118°F]
as tested to the Product Safety
Certification standards in the console
instructions for use.
Ingress
Protection:
IPX0 Ordinary Equipment
Environmental
Conditions:
Operation
Storage and
Transportation
Temperature
Limitation:
27 °C
10 °C
50 °C
-20 °C
Humidity Limitation:
85 %
10 %
90 %
10 %
Atmospheric
Pressure Limitation:
106 kPa
70 kPa
106 kPa
50 kPa