EasyManuals Logo
Home>Tektronix>Test Equipment>2467B

Tektronix 2467B Service Manual

Tektronix 2467B
536 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #163 background imageLoading...
Page #163 background image
Performance Check—2465B/2467B Service
i. Disconnect the test setup.
8. Check Channel Isolation.
a. Set:
CH 1, 2, 3 and 4
VERTICAL MODE
CHOP/ALT
CH 1 and CH 2
Input Coupling
CH 1, CH 2
VOLTS/DIV
CH 3, CH 4
VOLTS/DIV
TRIGGER SOURCE
A SEC/DIV
On
ALT
50 0 DC
0.1 V
0.1 V
CH 1
20 ns (knob in)
b. Connect the Primary Leveled Sine-Wave Generator
to the CH 1 OR X input connector via a 50
Q
BNC cable.
c. Set the generator frequency to 100 MHz and adjust
the output level for an 8-division display.
I. Replace the Primary Leveled Sine-Wave Generator
with the Secondary Leveled Sine-Wave Generator (with
the leveling head) and connect the generator to the
CH 1 OR X input connector.
m. Set the TRIGGER SOURCE to CH 1.
n. Set the generator output frequency to 400 MHz and
the output level for an 8-division display.
o. CHECK—Amplitude of each trace other than CH 1 is
0.16 division or less (discount trace width).
p. Move the signal to the CH 2 input connector and set
the TRIGGER SOURCE to CH 2.
q.
CHECK—Amplitude of each trace other than CH 2 is
0.16 division or less (discount trace width).
r. Disconnect the test setup.
d.
CHECK—Amplitude of each trace other than CH 1 is
0.08 division or less (discount trace width).
e. Move the signal to the CH 2 input connector and
change the TRIGGER SOURCE to CH 2.
f. CHECK—Amplitude of each trace other than CH 2 is
0.08 division or less (discount trace width).
9.
Set CH 1 and CH 2 DC Balance.
NOTE
For an accurate DC Balance setting, the instrument
MUST be allowed to warm up for 20 minutes before
performing the following steps.
a. Press both the CH 1 and CH 2 upper Input Coupling
buttons for approximately 1 second, then release them.
g.
Add a 50 Q BNC termination to the BNC cable and
move the signal to CH 3.
h. Set the TRIGGER SOURCE to CH 3 and adjust the
generator output for a signal display amplitude of 8
divisions.
i. CHECK—Amplitude of each trace other than CH 3 is
0.16 division or less (discount trace width).
j.
Move the signal to CH 4 input connector and set
TRIGGER SOURCE to CH 4.
b. VERIFY—DC BALANCE IN PROGRESS in top line
of readout. A flashing dot is also displayed. The display
returns to normal in approximately 15 seconds.
c. VERIFY—There is less than 0.2 division + 0.5 mV
vertical trace shift between adjacent settings of the CH 1
and CH 2 VOLTS/DIV as they are rotated through each of
their positions.
d.
VERIFY—There is less than 0.2 division vertical
trace shift between the CH 3 and CH 4 VOLTS/DIV
settings.
k. CHECK—Amplitude of each trace other than CH 4 is
0.16 division or less (discount trace width).
e. VERIFY—There is less than 1.0 division vertical
trace shift as the CH 1 and CH 2 VOLTS/DIV VAR
con-
trols are rotated fully CCW.
4-13

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix 2467B and is the answer not in the manual?

Tektronix 2467B Specifications

General IconGeneral
BrandTektronix
Model2467B
CategoryTest Equipment
LanguageEnglish

Related product manuals