EasyManuals Logo

Tektronix 4 Series Instructions

Tektronix 4 Series
22 pages
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #22 background image
How to sanitize a nonfunctional instrument
9.
Reassemble the instrument by using the disassembly steps in reverse order. Tighten the Torx T-10 screws to 0.65
Newton meters.
10.
Package the reassembled instrument, minus the Processor board, and ship to your nearest Tektronix Service Center
for repairs. A new Processor board will be installed
Repair charges
Replacement of damaged and missing hardware is charged according to the rate at the time of replacement.
14 4 Series MSO Opt. 4–SEC Declassication and Security Instructions