USER'S
CALIBRATION
General
The
following
procedure provides calibration infor-
mation
for
external adjustments
of
the 454A.
See
the com-
plete Instruction Manual
for
detailed calibration instruc-
tions.
Astigm~tism
Obtain a display
of
the calibrator waveform.
Adjust
ASTIG in conjunction
with
the FOCUS control for the best
definition
of
the square-wave display.
Trace Rotation
Obtain a free-running trace. Adjust TRACE
ROTATION
so
the trace aligns
with
the horizontal lines
of
the graticu
Ie.
Vertical Gain
Connect the 1
volt
calibrator
signal
to
the
input
connec-
tor
o.f
either vertical
input
channel. Set the vertical MODE
switch
to
display
that
channel and set
that
channel's deflec-
tion factor
to
.2 V/div. Adjust the front-panel
GAIN
adjust-
ment
of
the channel
for
exactly 5 divisions
of
deflection.
The best measurement accuracy when using probes
is
provided
if
the
GAIN
adjustment
is
made
with
the probes
installed.
®
454A
Step Attenuator Balance
Set both
Input
Coupling switches
to
GI\JD
and obtain a
free-running alternate dual-trace display.
Adjust
the STEP
ATTEN
BAL
control
in
each
input
channel
for
no trace
shift
when switching the appropriate attenuator switch
be-
tween the 2
mV
/div
and 10
mV
/div positions.
X Gain
(X-Y
Operation)
Set the
HORIZ
DISPLAY
switch
to
X-V and the
INT
TRIGGER switch
to
CH
1
ONLY.
Set the
CH
1 attenuator
switch
to
.2
V
/div
and the
CH
1 I
nput
Coupling switch
to
DC.
Apply
the 1
volt
calibrator signal
to
the
CH
1
Input
BNC
connector.
Adjust
the X GAII\J (X-Y) adjustment
for
exactly five divisions
of
horizontal deflection between the
displayed dots.
Horizontal Gain
Connect the calibrator signal
to
the
input
of
either verti-
cal
channel
and
set the vertical MODE switch
to
display this
channel. Set the HOR
IZ
DISPLAY
switch
to
display either
time
base.
Adjust
the appropriate triggering controls
for
a
triggered display at a
sweep
rate
of
one millisecond/
division. Check the display
for
one cycle
of
the calibrator
waveform
for
each
horizontal division.
For
more precise
timing measurements, refer
to
the calibration procedure
given in the Instruction Manual.
17