Specifications
Table 2: Digital channel input specifications, MSO2000 only (cont.)
Characteristic Description
Digital Record Length
1 Million Samples at all time base settings when a single set of 8 inputs are used, all from
the same physical half of the digital probe cable
1 Million Samples at time base settings from 100 sec/div to 200 μs/div when inputs from
both halves of the digital probe cable are used
500,000 Samples at time base settings from 100 μs/div to 2 ns/div when inputs from both
halves of the digital probe cable are used
Table 3: H
orizontal and acquisition system specifications
Characteristic Description
Long-term sample rate and
horizontal position time accuracy
±25 ppm o
ver any ≥1mstimeinterval
The limits below are for signals having amplitude >5 divisions, slew rate at the measurement
points of >2.0 d ivisions/ns, and acquired at >10 mV/div:
Condition
Time Measurement Accuracy
Single shot, full bandwidth selected ±[1 Sample Interval + 25 ppm × |reading|
+0.6ns]
>16 averages, full bandwidth selected ±[1 Sample Interval + 25 ppm × |reading|
+0.4ns]
Delta time measurement accuracy
Note: The Sample Interval is the time between the s amples in the waveform record
Instrument Range
DPO2024,
MSO2024
2 ns/div to 100 sec/div in a 1-2-4 sequence
Seconds/Division range
DPO2014,
MSO2014,
DPO2012,
MSO2012
4 ns/div to 100 sec/div
The minimum single pulse widths for guaranteed 50% or greater amplitude capture:
Instrument Minimum pulse width
DPO2024,
MSO2024
3.5 ns
FilterVu Peak Detect data record
pulse response
DPO2014,
MSO2014,
DPO2012,
MSO2012
7ns
Sample-rate 1 GS/s
Waveform Interpolation Only (sin x)/x interpolation i s provided
Record length 1 Million or 100,000 samples per record, user selectable
Waveform update r ate Minimum triggered acquisition rate is 5,000 wfm/sec