Installation
The oscilloscope displays a message when
the calibration is complete. Push Menu Off to
remove the me ssage.
7. After calibration, verify that t he status i ndicator
on the lower-bezel menu displays Pass.
Utility Page
Calibration
Signal Path
Pass
Factory
Pass
If it does no t, then recalibrate the oscilloscope
or have the oscilloscope serviced by qualied
service personnel.
Service personnel use the factory calibration
functions to calibrate the internal voltage
references of the oscilloscope u sing external
sources. Refer to your Tektronix eld ofce
or representative for assistance with factory
calibration.
NOTE. Signal Path Compensation does not include calibration to the probe tip. (See page 21, Compensating a P assive Voltage
Probe.)
DPO2000 and MSO2000 Series Oscilloscopes User Man ual 31